          恒溫(wen)恒濕(shi)箱是(shi)什麽原(yuan)因(yin)不(bu)制冷(leng)?
          點擊(ji)次(ci)數:2750 更新時(shi)間:2014-08-19

          原因(yin)壹(yi):
          1.由(you)於(yu)是(shi)溫度保(bao)持(chi)不(bu)住(zhu),觀(guan)察制冷(leng)壓縮(suo)機(ji)在(zai)試(shi)驗箱運(yun)行過(guo)程中是(shi)否能夠啟(qi)動(dong),壓縮(suo)機(ji)在(zai)環境試(shi)驗設備運行過(guo)程中都(dou)能夠啟(qi)動(dong),說(shuo)明從主電(dian)源(yuan)到(dao)各(ge)壓縮(suo)機(ji)的電(dian)器(qi)線路正常(chang),電(dian)器(qi)系統(tong)方面也沒(mei)有(you)問(wen)題(ti)。
          2.電氣系統(tong)沒(mei)有(you)問(wen)題(ti),繼續(xu)檢查(zha)制冷(leng)系統(tong)。首先(xian)檢(jian)查(zha)兩組制冷(leng)機(ji)組的低(di)溫(R23)級(ji)壓縮(suo)機(ji)的排(pai)氣(qi)和(he)吸氣壓力都(dou)較正常(chang)值偏低(di),而且(qie)吸氣壓力呈(cheng)抽(chou)空(kong)狀態(tai),說(shuo)明主(zhu)制冷(leng)機(ji)組的制冷(leng)劑量(liang)不(bu)足(zu)。
          3.用(yong)手(shou)摸(mo)主(zhu)機(ji)組R23壓縮(suo)機(ji)的排(pai)氣(qi)和(he)吸氣管路,發現(xian)排(pai)氣(qi)管路的溫(wen)度不(bu)高,吸氣管路的溫(wen)度也不(bu)低(di)(未結(jie)霜(shuang)),這(zhe)也說(shuo)明了(le)主(zhu)機(ji)組的R23制冷(leng)劑缺(que)乏。
          原因(yin)二(er):
          1.未確(que)定故(gu)障(zhang)原因(yin),結(jie)合(he)試(shi)驗箱的(de)控(kong)制過(guo)程進(jin)壹步(bu)確(que)認故(gu)障(zhang)原因(yin),該(gai)試(shi)驗箱擁(yong)有兩套制冷(leng)機(ji)組。
          2.壹為主(zhu)機(ji)組,另(ling)壹為輔助(zhu)機(ji)組,在(zai)降溫速(su)率(lv)較大時(shi),兩組機(ji)組同時(shi)工(gong)作(zuo),在(zai)溫度保(bao)持(chi)階(jie)段(duan)初期,兩組機(ji)組依然同(tong)時(shi)工(gong)作(zuo)。待(dai)溫(wen)度初步(bu)穩定下來(lai),輔助(zhu)機(ji)組停(ting)止(zhi)工(gong)作(zuo),由(you)主機(ji)組來維(wei)持溫(wen)度的(de)穩(wen)定。如(ru)果(guo)主(zhu)機(ji)組R23泄(xie)露(lu),會(hui)使(shi)主(zhu)機(ji)組的制冷(leng)效果(guo)不(bu)大,由(you)於(yu)降溫(wen)過(guo)程中,兩機(ji)組同時(shi)工(gong)作(zuo),故(gu)沒(mei)有(you)溫(wen)度穩(wen)定不(bu)住(zhu)的現象,而指示降溫(wen)速率(lv)降(jiang)低(di)。在(zai)溫度保(bao)持(chi)階(jie)段(duan),壹旦輔助(zhu)機(ji)組停(ting)止(zhi)工(gong)作(zuo),主機(ji)組又(you)無(wu)制冷(leng)作(zuo)用,試(shi)驗箱內的空(kong)氣就會緩(huan)慢上升(sheng),當(dang)溫(wen)度上升(sheng)到(dao)壹定程度,控(kong)制系統(tong)就會啟(qi)動(dong)輔助(zhu)機(ji)組來降(jiang)溫(wen),將(jiang)溫(wen)度下(xia)降(jiang)至設定值(-55℃)附近,然(ran)後(hou)輔助(zhu)機(ji)組又(you)停(ting)止(zhi)工(gong)作(zuo),如(ru)此反復,便(bian)會(hui)出現(xian)如(ru)圖3所示的故(gu)障(zhang)現象(xiang)。
          至此,已確(que)認生(sheng)產故(gu)障(zhang)的原(yuan)因(yin)是(shi)主機(ji)組的低(di)溫(R23)級(ji)機(ji)組的制冷(leng)劑R23泄(xie)漏(lou)。對(dui)制冷(leng)系統(tong)進(jin)行查(zha)漏,用(yong)檢漏(lou)儀和(he)肥(fei)皂水相(xiang)結(jie)合(he)的(de)方法(fa)檢查(zha),發現(xian)壹熱氣旁通(tong)電(dian)磁(ci)閥(fa)的閥(fa)桿裂了(le)約(yue)1cm的(de)細(xi)縫。更換此電磁(ci)閥(fa),對系統(tong)重新充(chong)氟(fu),系統(tong)運行正(zheng)常(chang)。由(you)於(yu)上文(wen)可(ke)以看出,對(dui)該(gai)故(gu)障(zhang)現象(xiang)的分(fen)析(xi)和判(pan)斷基本上是(shi)有易至難,先(xian)“外"後“裏",先(xian)“電(dian)氣(qi)"後(hou)“制冷(leng)"的脈(mai)絡進(jin)行分(fen)析(xi)和判(pan)斷的,熟(shu)悉和了(le)解(jie)試(shi)驗箱的(de)原(yuan)理和工(gong)作(zuo)過(guo)程是(shi)分析(xi)故(gu)障(zhang)判斷故(gu)障(zhang)的基礎。
           